GLAST The Gamma Ray Large Area Space Telescope

T&DF System Test Group

Convert Thermistors and RTD Counts to Resistance

RTD conversion where R = Resistance in ohms and c = ADC counts:
R = 1820*[ (c/4096 + 1/3)/(2/3 - c/4096) ]

Thermistor conversion where R = Resistance in ohms and c = ADC counts:
R = 4096*100000/c - 102000