Calorimeter calibration data.
A.Chekhtman.
May,10,2002
There is Eric
Grove’s presentation on calorimeter calibration at GLAST software meeting
January, 2001
Data used by actual calorimeter reconstruction code for
BFEM and BTEM.
- Muon
calibration: muon peak position (in ADC units) for each crystal face
(2*8*12=192 float values per tower)
- Charge
injection calibration:
- peak positions (in ADC units) for
different injected charge (up to 50 different DAC settings,
0<DAC<2000).
- For
each crystal face and each range calibration gives a table of 50 pairs
(DAC, ADC) => total data volume 2*50*4*2*12*8=76800 float values per
tower.
- Pedestals:
peak position (in ADC units) corresponding to 0 energy deposition in a
crystal. This information is in fact included in charge injection
calibration as data for DAC=0. Actually, pedestals are used by cal
reconstruction only for zero suppression purposes.
- Mu
slopes: coefficient converting light asymmetry into longitudinal position.
One float value for each crystal and each diode size. Total volume:
2*12*8=192 float values per tower.
Future modifications:
·
Muon calibration will be replaced by cosmic ray
calibration
o
Calibration
using ~10 different nuclei (H,
He, C,N,O,Si,Ne,Fe) =>there will be 10 peak positions for each crystal face
(1920 float values per tower)
·
One mu slope per crystal (and diode size) will be
replaced by complete calibration table (light asymmetry vs position) containing
12 pairs of float values => total
data volume 12*192=2304 float values.